A.H. Systems SAS-563B Manuale operativo - Pagina 7

Sfoglia online o scarica il pdf Manuale operativo per Antenna A.H. Systems SAS-563B. A.H. Systems SAS-563B 13. Active loop antenna
Anche per A.H. Systems SAS-563B: Manuale operativo (14 pagine)

A.H. Systems SAS-563B Manuale operativo
A.H. Systems Model SAS-563B Active Loop Antenna
Each antenna is individually calibrated per IEEE 291-1991 and comes with a
calibration certificate references the NIST traceable test equipment. Initially
the antenna comes with a 1 year calibration interval, however depending on
the type of use and standards that the antenna is being used, this interval can
be adjusted.
The calibration tables shown provide a listing of the frequency of operation
and its correction factor in dB. The field strength is the receiver voltage in
dBV plus the correction factor (refer to the correction factor calibration) plus
any cable loss. When making a measurement, mount the antenna on an
appropriate mast or tripod with the plane of the loop antenna facing the DUT.
SHIELDING EFFECTIVNESS TESTING
Both the SAS-563B and the SAS-564 loop antenna is used for shielding
effectiveness testing. The SAS-563B is used as the receive antenna and
should be connect to an analyzer or receiver, where the SAS-564 Loop
antenna is a passive loop antenna used to transmit.
© A.H. Systems inc. – Mar 2021
REV E
7