A.H. Systems Model SAS-563B Active Loop Antenna
Each antenna is individually calibrated per IEEE 291-1991 and comes with a
calibration certificate references the NIST traceable test equipment. Initially
the antenna comes with a 1 year calibration interval, however depending on
the type of use and standards that the antenna is being used, this interval can
be adjusted.
The calibration tables shown provide a listing of the frequency of operation
and its correction factor in dB. The field strength is the receiver voltage in
dBV plus the correction factor (refer to the correction factor calibration) plus
any cable loss. When making a measurement, mount the antenna on an
appropriate mast or tripod with the plane of the loop antenna facing the DUT.
SHIELDING EFFECTIVNESS TESTING
Both the SAS-563B and the SAS-564 loop antenna is used for shielding
effectiveness testing. The SAS-563B is used as the receive antenna and
should be connect to an analyzer or receiver, where the SAS-564 Loop
antenna is a passive loop antenna used to transmit.