Filter Replacement
Cold Evaporation Filter Replacement
The cold evaporation filter should be replaced every eight months, or whenever indicated
by the yellow 1 illuminated on the Filter Replacement Indicator on the Control Panel.
1.
Remove the top cover from the back of the device.
2.
Grasp the handle of the HOCl generator and turn it to the right to release it from
the base.
3.
Remove the HOCl generator.
NOTE: Handle the Cold evaporation filter by the end caps to avoid touching
the Cold evaporation filter and preventing possible contamination of the
filter element.
4.
Unscrew the cold evaporation filter from the generator and discard.
5.
Remove the replacement cold evaporation filter from packaging and screw it into
the inner bottom of the HOCl generator.
6.
Replace the HOCl generator in the Sterilizer and turn the handle to the left to lock it
into place on the base.
7.
Turn on the Sterilizer. After replacing the filter, press and hold the Filter Replacement
Indicator for 3 seconds to reset the filter life countdown.
NOTE: If you replace the filter before the accumulated life time is reached, the
filter life timer cannot be reset.
9
H13 True HEPA Filter Replacement
The filter should be replaced every 12 months or whenever indicated by the red 2 illuminated
on the Filter Replacement Indicator on the Control Panel. However this interval is intended
as a guideline only. In environments with a high concentration of contaminants such as dust,
pet dander, or cigarette/cigar smoke you may need to replace the filter more frequently.
NOTE: The HEPA filter is not washable. DO NOT immerse it in water.
1.
Make sure the device is OFF and unplugged.
NOTE: If you have difficulty removing the bottom cover, remove the top cover
prior to removing bottom cover.
2.
Detach and remove the magnetic bottom cover from the unit.
3.
Remove using the pull tabs and replace the replacement filter into the device.
NOTE: Ensure that the pull tabs are facing outward to facilitate filter removal later.
4.
Inspect the screen on the bottom cover and if necessary clean any excess dust or dirt
from the screen using a hand held vacuum.
5.
Replace the bottom cover by placing the bottom cover into the slots on the device.
Gently push the top of the bottom cover into place.
6.
Turn on and hold the Filter Replacement Indicatot for 3 seconds to reset the filter life
countdown.