DiaTecne PulsePen WPP001-ETT Manuale d'uso - Pagina 8
Sfoglia online o scarica il pdf Manuale d'uso per Apparecchiature mediche DiaTecne PulsePen WPP001-ETT. DiaTecne PulsePen WPP001-ETT 16.
Maintenance and Cleaning
No particular periodic maintenance or calibration operations are required on the instrument.
⚠ In the case of prolonged non-use, remove batteries.
⚠ Before use, the metallic disc of the wTn1 probe, the patient cable CV010 and the probe's case must be
cleaned, using a clean, dry cloth or dampened with a small quantity of alcohol.
⚠ Be very careful to keep the alcohol or other liquids from penetrating into the wTn1 probe or into the
other units because this could cause serious problems, irreparably damaging internal parts.
⚠ The CV010 Ecg cables are thin and flexible in order to be easily handled. One must avoid pulling or
bending them at sharp angle so as not to damage them.
Warnings and Precautions for Use
⚠ It is very important to read the following items before the use of the device. Improper use may cause
very severe consequences.
• This equipment is intended to be used only by medical/paramedical personnel in medical environment.
• Do not use the equipment in operating room, or in any case where there are inflammable gases/
substances.
• Do not use the equipment for invasive or intra-cardiac applications or at direct contact with internal body
parts.
• Data elaboration must be performed by the doctor through dedicated software.
• Do not sterilize neither in autoclave nor with liquid substances.
• Avoid mechanical shocks to the pressure plunger of the wTn1 probe, like collisions or falls.
• Regularly clean the disc of the wTn1 probe, the terminals of the patient cable CV010 and the case after
each use, as indicated in the "Maintenance and Cleaning".
• Keep the units wTn1 and wEc1, with its patient cable, at a distance of no less than 1.5 meters from the
computer and the same computer at no less than 1.5 meters from the patient.
• Avoid touching simultaneously any part of the computer, including the wRs1 unit, and one or both of the
wEc1, including the patient cable, and the wTn1 units.
• Do not immerse any part of the device into water or other substances, or expose it to sprays. Absolutely
never use Gel on the pressure sensor of wTn1.
• Do not open the device for any kind of maintenance work; in the case of malfunction of the device, contact
DiaTecne s.r.l.
• In the case of any abrasion, sheath damage, or any kind of defects appears in the patient cable CV010,
immediately suspend use of the device and contact Diatecne s.r.l. for repair/replacement.
• Do not use the device if breakage occurs to any part, do not try to repair it but contact Diatecne s.r.l. for
repair/replacement.
• Do not make changes of any kind to the device and never use accessories other than those provided.
• Keep the tonometric probe wTn1 and patient cable CV010 terminals away from electrical outlets/sockets
and surfaces where there may be potentially dangerous voltages.
• Use a battery-powered (portable) computer or, alternatively, an AC-powered computer, compliant with the
current medical regulations.
• Use the device at a safe distance from sources of electromagnetic disturbance such as cordless telephones
operating at radio frequency/cellular phones, Bluetooth and WiFi devices or other equipments emitting
high frequency electromagnetic waves.
• During examination, keep the wEc1 unit at a distance of at least 20 cm away from both the patient and
operator, limit the duration of contact of the wTn1 unit with both patient and operator for the time
! 8
!
User Manual PulsePen - V. 4.6