SETTING UP THE METER

Before using your meter for the first time or
whenever replacing its batteries, check and
adjust the following settings:
Open the setup mode (a)
The meter is switched off (a test strip is not
inserted).Press the button marked S beside
the batteries.
1. Setting the date
The sequence for setting the date is
YEAR ➞ MONTH ➞ DAY. When YEAR/
MOTH/DAY are flashing consecutively,
press the button marked M to select the
correct setting. Next, press S.
2. Setting the date format
Press the main button M on the front of
the meter to select the required date
format (12 or 24 hour). Then press S.
3. Setting the time
When HOUR/MINUTE are flashing
consecutively, press the button marked M
until the correct time is displayed. Then
press S.
4. Turning sound effects on/off
For sound effects press the button marked
M on the display to switch between "On"
and "Off" Then press S.
5. To set a reminder
Your meter has four reminder settings.
The meter displays "ON" or "OFF" and
If you do not want to set a reminder
notification, then press S to skip this step.
Otherwise press the button marked M to
select "ON" and then press S.
10
When the hour/minute are flashing, press
M to select the desired hour/minutes.
Then press S and continue to the next
alarm setup.
PLEASE NOTE: When the reminder
notification goes off, press M to switch it
off. The reminder will sound for 2 minutes
and then switch off automatically.
6. Turning on Bluetooth
(Bluetooth® is optional)
When "bt" is visible on the display press
M and select "ON" or "OFF". Then press S.
PLEASE NOTE: This means that data is
transferred using Bluetooth®.
If you select "ON", then the results are
transmitted immediately after the test.
NOTE:
• ALL of these parameters can be changed
in the setup mode.
• If the meter is not used for a period of
3 minutes while in setup mode,
it switches off automatically.
.
Measuring methods:

for blood sugar testing

You can measure your blood sugar levels
in three different ways: General, AC (before
meals) and PC (after meals). You can switch
between the different modes as follows:
1. Start with the meter switched off.
Insert a blood sugar test strip into the slot
to turn the meter on. The display shows ,
"
" flashing " " and "GLU".
2. Press M to switch between General,
AC and PC.
For β-ketone testing
3. There is only one option for measuring
your β-ketone levels: Gen. Start with the
meter switched off. Insert a β-ketone test
strip into the slot to turn the meter on. The
display shows "
", a flashing" ", "Gen"
and "KET".

QUALITY CONTROL TEST

When should you test the device
using control solution?
You do not need to perform a test with
the control solution when first using the
Diatesse XPER.
• When you think that either the meter or
the test strips are defective.
• When your test results do not match the
way you are feeling, or when you believe
the results are not accurate.
• When performing a test.
• If you have dropped the meter or have
reason to believe it is damaged.
Test strips (c), control solution (d), lancing
device (e) or sterile lancets (f) might not be
included in the kit (please check the list of
contents on your product's packaging). These
articles can be ordered separately. Ensure that
you have a sufficient supply of all the articles
you require for performing tests.
Performing a test with
the control solution.
In order to perform a test with the control
solution, you require (b), (c) and (d).
1. Insert the test strip into the slot to turn
the meter on.
Wait until the meter displays"
" and
a flashing " ".
2. Apply the control solution (g)
Shake the bottle of control solution well
before use. Squeeze out the first droplet
and wipe it off. Squeeze another droplet
onto the point of the bottle's cap. Hold the
meter so that the absorbent opening of
the test strip comes into contact with
the drop and absorbs the control solution.
Once the control window contains
sufficient control solution, the meter starts
an automatic countdown.
11