While holding the knurled outer barrel stationary with
the soft-jaw pliers, use a suitable slotted screwdriver to
unscrew and remove the slotted end-screw from the
knurled outer barrel (see
Figure 5 – Loosen and remove the slotted end-screw
After the slotted end-screw has been removed, rotate
the knurled outer barrel clockwise (if necessary) until
the smooth end of the knurled outer barrel contacts the
stationary inner barrel (see
Figure 6 – Snug the knurled outer barrel down
The next step is to remove the e-clip from the groove in
the end of the plunger shaft. In order to expose the end
of the plunger shaft to access the e-clip, the plunger
must be held in the fully depressed position. One way
to do this is to use a dovetail mounting disk from an old
binocular head (see
Figure
Figure 7 – Mounting disk from an old binocular head
Teardown, Cleaning, and Reassembly of the Miscellaneous Parts of the Olympus BH-2 Microscope Frames
Figure
5).
Figure
6).
7) to depress the plunger.
Carefully place the circular dovetail of the mounting
disk (see
Figure
7) into the dovetail recess for the
viewing head, using the circular dovetail to depress the
plunger while seating the mounting disk into the recess
(see
Figure
8).
Figure 8 – Place mounting disk into dovetail recess
Once the mounting disk has been properly seated into
the dovetail recess for the viewing head, slide the
mounting disk towards the locking screw and observe
that the end of the plunger shaft, with the attached
e-clip, protrudes from the knurled outer barrel of the
thumbscrew when the mounting disk has been slid as
far as it can go towards the locking screw (see
Figure 9 – Slide the mounting disk to expose the e-clip
Use a suitable clamp with non-marring jaws to hold the
mounting disk in the slid-over position in the dovetail
recess, thereby keeping the e-clip on the end of the
plunger shaft fully exposed (see
clamp is secured to hold the mounting disk in this
position, the plunger will remain depressed such that
there is sufficient access to remove the e-clip from the
plunger shaft.
Figure
Figure
10). Once the
Revision 1
Page 10 of 45
9).