Phenomena
The nosepiece does not move.
Auto-focusing does not work.
Repair request
If you cannot improve the phenomena after taking the above remedies, please contact us for assistance.
At that time, please tell them the following information as well.
Product name and abbreviation (Example: BXC-CBB)
l
Product number
l
Phenomena
l
6. Troubleshooting
Cause
Wrong cable is used.
Wrong command is used.
The nosepiece setting to select
either the 6Position-NP or
5Position-NP on the DIP switch is
wrong.
Only a specific hole (partial
segment) is used.
The safety lock is not attached.
Remedy
Use a correct cable.
Refer to the command reference
manual.
Specify the correct nosepiece
setting on the DIP switch.
Refer to "To extend the life of the
nosepiece" and evenly apply the
grease.
Attach the short plug provided
with the product or a lock
mechanism.
Reference
page
23
10
31
34
6.3 Drive system