Olympus CK40M FRAME Manuale di riparazione - Pagina 11

Sfoglia online o scarica il pdf Manuale di riparazione per Microscope Olympus CK40M FRAME. Olympus CK40M FRAME 38. Compact inverted metallurgical microscope

Olympus CK40M FRAME Manuale di riparazione
CK40M
B. DISASSEMBLY AND ASSEMBLY PROCEDURES
(3)
Prism (LP142600) mounting position
M S
Surface A
CK40MKC02
B
TR body (AD064400)
Prism (LP142600)
Mount the prism positioning jig (CK40MKC02) on the TR body (and tighten it firmly). Set the
prism against the jig. Determining this as the prism fixing position, apply silicone adhesive
(OT1873) at least 15 mm on the both sides of the prism.
Positioning the surface A as the bottom, insert the M4 screw (AB453100) used for CK40 to
location B shown in the above illustration to fix the prism while preventing its falling. It takes
about 12 hours for OT1873 to go solid because it is silicone adhesive.
B-4