Olympus CK40M FRAME 수리 매뉴얼 - 페이지 11
{카테고리_이름} Olympus CK40M FRAME에 대한 수리 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Olympus CK40M FRAME 38 페이지. Compact inverted metallurgical microscope
CK40M
B. DISASSEMBLY AND ASSEMBLY PROCEDURES
(3)
Prism (LP142600) mounting position
M S
Surface A
CK40MKC02
B
TR body (AD064400)
Prism (LP142600)
Mount the prism positioning jig (CK40MKC02) on the TR body (and tighten it firmly). Set the
prism against the jig. Determining this as the prism fixing position, apply silicone adhesive
(OT1873) at least 15 mm on the both sides of the prism.
Positioning the surface A as the bottom, insert the M4 screw (AB453100) used for CK40 to
location B shown in the above illustration to fix the prism while preventing its falling. It takes
about 12 hours for OT1873 to go solid because it is silicone adhesive.
B-4