CX21 MAINTENANCE MANUAL

2. Mechanical Adjustment

2-1 Preparation for the tension adjustment of X/Y-wire

If a specimen image is moved when the stage is brought into the desired position of
specimen, it is necessary to adjust the wire tension of stage.
1)
Take off the mechanical stage (A) from the plane stage (B)
by removing the two screws (*1).
(A)
2)
Remove the specimen holder (C) from the
inner guide (D).
Screw: 3PSK2X4SA (*2), 3pcs.
X-wire
* For the tension adjustment of X-wire,
refer to next page.
(B)
(a)
*1
*2
(C)
(B)
arrow view (a): bottom
*1: Fixing screws of mechanical stage
Screw: AB3X10SA (*1), 2 pcs.
3)
Remove the Y-cover (E).
Screw: CUTB3X8SA (*3), 3pcs.
Y-wire
(D)
(E)
*3
* For the tension adjustment of Y-wire,
refer to next page.
- 14 -
REPAIR PROCEDURE
*1
(A)
<indicated by arrows>