Placing the Specimen
Put the specimen in the glass-bottom dish (with cover glass thickness
of 0.16 to 0.18 mm) and place the dish on the center of the stage.
} If the dish is prone to slide on the stage, attach the stage clips (IX-SCL)
1 or use the provided weight.
5
Setting the Focus Drift Compensation Function
1. Run the "Application Software for PC."
2. Set up the AF (autofocusing) functions.
The following parameters should be set to start AF.
In one-shot mode
· Offset : Setup of the amount of compensation between the cover glass
top surface and target position.
· AF zone : AF search zone (AF center, width and range)
In continuous focus mode
Confirm that the cover glass top surface is detected in the one-shot
mode before switching to the continuous focus mode.
} For details, refer to the instruction manual and Help of the application.
When using the MetaMorph application software, install IX2-
CAUTION
BSW Ver. 1.7 (in the CD provided with this product) in your PC
in advance.
IX81-ZDC2
(Fig. 3)
10