Olympus SZX-RFL2 Manuale di istruzioni - Pagina 8
Sfoglia online o scarica il pdf Manuale di istruzioni per Apparecchiature di laboratorio Olympus SZX-RFL2. Olympus SZX-RFL2 22. Coaxial fluorescence attachment
![Olympus SZX-RFL2 Manuale di istruzioni](/uploads/products/server2/4291226/4291226.png)
3-2 Detailed Assembly Procedures
³
Fig. 1
@
Fig. 2
³
Fig. 3
Attaching the Coaxial
1
Fluorescence Illuminator
1. Remove the observation tube from the microscope.
@
2. Attach the coaxial fluorescence illuminator onto the microscope body by
aligning the illuminator's positioning groove @ with the microscope's
²
positioning pin ².
3. Tighten the observation tube clamping screw ³ using an Allen screw-
driver.
4. Attach the observation tube, which has been removed in step 1, above the
coaxial fluorescence illuminator.
2
Installing the Mirror Units
# Handle the filters and mirrors of the mirror units so as not to stain
them with fingerprints, etc.
When it is required to change the initial setup later, be always sure to
set the shutter slider of the illuminator tube in the fully pulled-out
(light-excluding) position or set the main switch of the power supply
unit to "
# Up to 3 mirror units can be installed by numbering them 1, 2 to 3 from
rear to front (note that this numbering order is the reverse of the
indications on the switch knob).
1. Loosen and remove the lock knob ² of the cover panel @ on the illumi-
nator tube. The cover plate is attached by magnetism on the illuminator
tube, but can be removed easily by pulling lightly.
2. Slide the mirror unit switch knob ³ to a position in which each mirror unit
can be installed easily.
3. Using the Allen screwdriver provided with the microscope body, slightly
²
loosen the mount dovetail lock screw |.
4. Screw in the lock knob ² into the mirror unit ƒ to be installed, insert the
mirror unit straight and all the way into the illuminator tube along the
mount dovetail, and tighten the mount dovetail lock screw | securely.
# If the mirror unit is inclined vertically or horizontally, smooth insertion
is not possible.
|
ƒ
²
" (OFF) in advance.
SZX–RFL2
(Fig. 1)
(Figs. 2-4)
4