Olympus SZX-RFL2 Manuale di istruzioni - Pagina 9
Sfoglia online o scarica il pdf Manuale di istruzioni per Apparecchiature di laboratorio Olympus SZX-RFL2. Olympus SZX-RFL2 22. Coaxial fluorescence attachment
![Olympus SZX-RFL2 Manuale di istruzioni](/uploads/products/server2/4291226/4291226.png)
...
Fig. 4
}The SZX-STF Focusing Adapter for Ring Mount makes it possible to build a system using the SZX-ZB7 (SZX7
microscope frame) + SZX-RFL2 (coaxial reflected fluorescent light illuminator) + SZX-ST/SZX-ILL series (SZX base
series).
The use of the SZX-STF improves the stability of the microscope and reduces oscillations, which pose a problem
during fluorescent light imaging under extended exposure or during high-magnification observations.
1. External View and Nomenclature of the SZX-STF Adapter
Microscope clamping screw
Also used to clamp the LSGA
mount.
LSGA mount
Uses the same clamping
screw holes (x 3)
2. Attaching the SZX-STF Adapter
³
ƒ
|
5
When Using the SZX7 Microscope Frame
Microscope mount hole
76 mm
@
²
5. Attach the mirror unit display plate ... provided with the mirror unit to one of
the display attaching positions on the front panel by means of magne-
tism.
}Install other mirror units in the same way as detailed above.
# One or two mirror unit dovetails may not have a mirror unit installed.
Even in this case, tighten their lock screws tightly.
If the lock screws are not tightened, they may slip out during the use
of the unit, making it impossible to slide the mirror unit switch knob.
6. After having installed the required mirror units, place the cover panel @ in
the original position by tightening the lock knob ².
1. Remove the cap @ of the focus adjustment section by inserting the
tip of a pointed object into the notch.
2. Using the Allen wrench provided with the SZX7 microscope frame,
turn the dovetail clamping screw inside the focusing module cap by
2 or 3 turns (counterclockwise) to loosen the screw.
3. Fit the mount dovetail ³ on the rear of the adapter gently into the
mount dovetail ² of the focusing section.
# Do not fit obliquely or push in by force, as this may damage the
functionality.
4. After fitting the dovetail till the stop position, tighten the clamping screw
and attach the cap again.
5. Fit the SZX7 microscope frame gently into the microscope mount
hole | and tighten the microscope clamping screw ƒ.
Mount dovetail