4. Tighten the two stop screws ... on the body mount firmly using the Allen
wrench (small).
5. Using the Allen screwdriver, tighten the tension screw † while lightly push-
ing it downward.
6. Rotate the zooming knob on the right side panel of zoom microscope
body toward the front, to the minimum zoom position.
7. Rotate the knob ‡ on the coupling all the way to the front until it is
stopped.
8. While pulling the knob ‡ lightly toward the left, tighten the stop screws Š
firmly using the Allen wrench (small).
8
Attaching the Filter Changer
}Attach the filter changer in the same way as attaching the observation
tube.
9
Attaching the Shield Plate
Align the mount hole ² on the light shield plate with the screw hole @ on
the lighting head, insert the provided screw and tighten it firmly using the
Allen screwdriver.
²
SZX–RFL3
(Fig. 9)
10