Olympus SZX-RFL3 Manuale di istruzioni - Pagina 14

Sfoglia online o scarica il pdf Manuale di istruzioni per Apparecchiature di laboratorio Olympus SZX-RFL3. Olympus SZX-RFL3 22. Fluorescence attachment

@
Fig. 12
²
Fig. 13
³
Fig. 14
|
ƒ
Fig. 15
Setting the Burner Hour Counter
1. Press the center section ² of the reset button @ on the power supply
unit's front panel to reset the hour counter to 000.0.
}The hour counter shows elapsed time in hours. For safety's sake, replace
the burner when the hour counter indicates 200 (USH102D) or 300
(HBO103W/2) hours.
Mercury Burner Replacement
1. In order not to impair the safety of the equipment, replace the burner
when it has been used for 200 (USH102D) or 300 (HBO103W/2) hours.
The burner may crack if used beyond the specified life time.
2. Before replacing the burner, wait at least 10 minutes or until the lamp
and lamp housing have cooled down after turning the burner off.
Before removing the burner, confirm that the main switch on the power
supply unit is "
the output connector on the power supply unit. Refer to page 11 for
details on replacement procedure.
3. After replacing the burner, reset the hour counter to 000.0 as outlined
above.
13
Setting the Power Supply Unit
Cables and cords cam easily be damaged when bent or twisted. Do
@
not subject them to excessive force.
Make sure that the main switch is set to "
the power cord.
²
Always use the power cord provided by Olympus. If no power cord is
provided, please select the proper power cord by referring to
"PROPER SELECTION OF THE POWER SUPPLY CORD" at the end of
this instruction manual.
1. Verify that the voltage and frequency of the input power supply match the
requirements indicated on the name plate @.
(100 V systems can be used with voltages in the 100 to 200 V range and
200 V systems can be used with voltage in the 220 to 240 V range, both
with frequencies of 50 to 60 Hz.
2. Securely plug the burner socket connection cord into the power supply
unit's connector ².
3. Plug the power supply unit's power cord into its power input connector ³,
then plug the power plug | into the wall power outlet ƒ.
Be sure to supply power from a grounded 3-conductor power outlet
using the proper power cord. If the power outlet is not grounded
properly, Olympus can no longer warrant the electrical safety perfor-
mance of the equipment.
" (OFF) and unplug the connecting cord plug from
" (OFF) before connecting
SZX–RFL3
(Figs. 14 & 15)
12