Accuris SmartDoc 2.0 Manuale di istruzioni - Pagina 6
Sfoglia online o scarica il pdf Manuale di istruzioni per Apparecchiature di laboratorio Accuris SmartDoc 2.0. Accuris SmartDoc 2.0 8. Imaging enclosure
Anche per Accuris SmartDoc 2.0: Manuale di istruzioni (8 pagine), Filtro e manuale d'uso (1 pagine)
Note: The SmartDoc uses a timing circuit, and the lights will automatically turn
off after 5 minutes. This timer prevents excess heat from building up from the
powerful LED lights.
5. GEL VIEWING AND BAND CUTTING
The SmartDoc Illumination Base can be used with the orange filter cover for
visualizing stained nucleic acids Gel Viewing separated on agarose or acylamide
gels.
Orange filter cover placed for gel viewing
1. Set the gel preparation platform onto a level lab bench or table.
2. Glean the glass surface with alcohol or glass cleaner and allow to dry.
3. Carefully place a stained electrophoresis gel onto the square glass
surface using care to eliminate bubbles under the gel.
4. Note: The maximum sized gel to fit onto the glass surface is 15x15cm,
cut away any overhanging edges of the gel if required.
5. Place the blue illumination base over the gel preparation platform,
taking care that it nests correctly into place.
6. SMARTDOC PHOTO FILTERS
E5001-ORANGE
(included)
A photo filter is required to capture quality images with the SmartDoc 2.0 system.
The orange photo filter, part E5001-ORANGE, is included with the system. This
filter blocks the visible blue wavelengths coming from the illumination base or
Orange filter cover set at an angle for gel
access and band cutting
E5001-535
(optional)
6
E5001-590
(optional)