14. Cleaning
Clean the Easy hygiene chair with detergent/disinfectants commonly used
in healthcare (alcohol-based). The pillars should be cleaned with agents that
dissolve grease and also remove lime stains. Raise the chair to maximum
height when cleaning the pillars. Apply the cleaning agent and leave it to
work according to the instructions (also allow it to run down inside the
pillars); flush out. NOTE: Also clean the pillar base plate periodically. Do
so by removing the bottom plug (see fig. 17) and applying cleaning agent
directly in the pillar. Allow to work; flush out.
Read the warning text on the packaging/cleaning agents and in their
data sheets.
NOTE: To avoid crush injuries, make sure protective covers (fig. 19)
are always put back
15. Troubleshooting when the chair is inoperable
NOTE: Chair cannot be operated above 65 cm seat height if the chair is
tilted forward, nor can the seat be tilted forward if seat height is greater than
65 cm.
Before troubleshooting begins, always restart the battery (see page
4. Hold in the battery's OFF button for approx. 5 seconds. Wait for
one minute. Activate the battery using the ON button).
1 Check the battery charge via
the remote control's battery
indicator (if necessary, activate
the remote control by pressing
any button).
1a Indicator shows green ==>
Calibrate the chair (See page
7 ).
1b Indicator shows red == >
Charge the battery
1c Indicator lamp out ==>
Connect the remote
control to the charger. If
charging is indicated in the
remote control, charge the
battery as per instructions. If
charging is not indicated,
proceed to 2.
If the chair remains in operable following completion of the above, contact
customer service at GATE.
8
EA SY HI GH BA CK
2 Switch off the battery manually
using the battery Off button
(see page 4). Wait one
minute. Activate the battery
by pressing its On button (see
page 4). Check that the battery
indicator lamp is lit or flashing
Indicator lamp lit/flashing == >
Replace remote control
and charge the battery as per
the instructions.
Indicator lamp not lit, not
flashing ==> Replace the
battery