Nikon NR-5500 Manuale di riparazione - Pagina 21
Sfoglia online o scarica il pdf Manuale di riparazione per Strumenti di misura Nikon NR-5500. Nikon NR-5500 25. Auto refractometer
TEST PROGRAM
p— [+000] — [+012] -> [+025] — [+037] -*••••
[+100]
-
0. OOD
+0. 12D
+0. 2 5 D
+0. 37D
+1. OOD
— [-012]
[-062] «- [-075] «- [-087] — [-100] --
-0. 1 2 D
-0. 62D
- 0 7 5 D
-0. 37D
-1. OOD
After establishing the shifting amount, turn DIP switch No. 5 located on the
right side of the Main body to the ON position and press the [SET] key.
The amount of parallel shifting will then be stored in the EEPROM and two short
beeps will be heard. Then return DIP switch No. 5 to the OFF position.
5 .RETRO mode (RETRO image brightness adjustment)
At RETRO mode, screen image brightness and anterior illumination intensity canbe
adjusted.
(1) Connect cables.
• Using co - axial cable tool
Connect the auxiliary monitor output (J3) on the CPU board and the CHANNEL 1
on a oscilloscope.
• Using oscilloscope's probe
Connect a probe to No.l pin at J3 connector.
Connect a ground cable to No.2 pin at J3 connector.
* Do not make a short circuit to prevent from a CPU damage.
(2) Oscilloscope setting
*Use a CHANNEL 1.
• Each setting as follows.
SEC(TIME)/DIV --*
2 ms or 5 ms.
VOLTS/DIV
—->
0. 2 V AC.
(D Preparation for NR - 5500
• Attach a front cover. Unit must have the dust protection glass.
• Mount a model eye on a forehead rest bar.
• Align a model eye correctly.
• Get a regular measuring mode and then set a RETRO mode with pressing the [RET]
key.
C6 - 1 5]
Repair Manual