Nikon NR-5500 Manuale di riparazione - Pagina 24
Sfoglia online o scarica il pdf Manuale di riparazione per Strumenti di misura Nikon NR-5500. Nikon NR-5500 25. Auto refractometer
PROTECT DATA SETTING
@ Find the [PRE SET] key displayed to the left of the [END] key in the system data
setup menu.
(D Press the [PRE SET] key on the screen, and the protect data setup menu wall appear.
Using the [ t ] and [ 1 ] keys, select the desired item and set the corresponding data
with the [ELSE] key.
* Description of Protect dada
(1) RS - 232C : Setting up the external communication mode.
=NONE : No communication (Select this when not connected)
=NNKI : Transmit data in new NK mode on request from external device.
=NNKE : Transmit data to external device in new NK mode on each end of
measurement (triggered by ''Print" switch).
=NK
: Transmit data in old NK mode (same as NR - 5000).
(2) IC CARD : Setting up the built - in IC card reader/writer
=NONE : No IC card reader/writer is built - in.
=NNK
: Program data into IC card in new NK mode.
=NK
: Program data into IC card in old NK mode.
C7 — 2]
Repair Manual