Camille Bauer KINAX N702-INOX HART Manuale di istruzioni per l'uso - Pagina 14
Sfoglia online o scarica il pdf Manuale di istruzioni per l'uso per Trasmettitore Camille Bauer KINAX N702-INOX HART. Camille Bauer KINAX N702-INOX HART 19. Via the hart field communicator 475
Anche per Camille Bauer KINAX N702-INOX HART: Manuale di istruzioni per l'uso (15 pagine), Manuale di istruzioni per la programmazione (15 pagine)
7. Electric connections
7.1 Inclination transmitter
The inclination sensor is connected via a highly fl exible,
shielded signal line with 2 strands and PUR steath. The same
is preassembled at the factory, included in the delivery and
available in different lenghts.
Permissible cable-Ø:
conductor cross-section:
Wiring connection:
White (wt) = +24V
Brown (br) = 4...20 mA
N
μC
S
PSU
wt
18...33V DC
• Do not electricly modify the device nor
carry out any wiring work when energised.
• Do not plug in or unplug electric connection
when energised.
7.2 HART-Signal
The HART signal is directly tapped via the 4 ... 20mA signal
line. The measuring circuit must have a minimum load
of 240Ω. Connect the HART communicator or the HART
modem via cable terminals or the existing communication
sockets of a power supply unit or separator. The connection
is independent of polarity.
USB
HART-USB
Modem
R
L
Transmitter-
power supply
unit
24V
8. Electric commissioning
•
Provide a separate power supply for the device in case
of consumer loads with high interference levels.
•
Install the entire plant in an EMC-compatible manner.
Installation environment and wiring can affect the EMC
of the device.
9. Maintenance
The device is free of maintenance. Repairs may only be
carried out by authorized authorities.
14
5.9 mm [0.232]
2 x 0.34 mm
2
HART
+
-
br
Load resistor
4...20mA
HART
Communicator
10. Programming
The inclination transmitter may be programmed via HART
modems (e.g. ifak system type isHRT USB) or via HART
communicator (e.g. EMERSON type 475). The cable termi-
nals of the HART modems or communicators are directly
connected to the 4 ...20mA signal line.
• A reset to factory settings is not possible.
• The inclination transmitter must be con-
nected according to the wiring diagram
• The smallest programmable angle is 10°.
• Programming instructions for HART
Field Communicator 475 available on
our homepage www.camillebauer.com.
Use case 1: Set start and end value
1. Setting zero point (Command 43: Set primary variable zero)
2. Put transmitter into initial position
program 0% = 4 mA (Command 45: Trim Loop Current
Zero)
3. Put transmitter into fi nal position
program 100% = 20 mA (Command 46: Trim Loop Current
Gain)
4. Defi nition of direction of rotation
0 Point (43)
cw / ccw
•
0% = 4 mA (37)
100% = 20 mA (36)
•
•
100% = 20 mA (36)
Use case 2: Set of fi xed measuring range
1. Set zero point (Command 43: Set primary variable zero)
2. Set measuring range
e.g. -15° = 4 mA ... +30° = 20 mA or
-30° = 4 mA ... +30° = 20 mA
Command 36: Primary Variable Upper Range Value
Command 37: Primary Variable Lower Range Value
11. Ordering information of variants
Description
KINAX N702-INOX, Cable length 1,5m
Set to 0 ... 60° (Special version)
KINAX N702-INOX, Cable length 5,0m
Set to -180 ... 179.99°
KINAX N702-INOX, Cable length 10,0m
Set to -180 ... 179.99°
KINAX N702-INOX, Cable length 30,0 m
Set to -180 ... 179.99°
Further cable lengths on request
12. Accessories
Description
Mounting plate N7xx INOX
(already included in delivery)
Kit mounting clamp N7xx INOX
(already included in delivery)
cw / ccw
0% = 4 mA (37)
•
•
0% = 4 mA (37)
100% = 20 mA (36)
Article No.
172 510
172 536
172 544
173 922
xxx xxx
Article No.
172 619
172 627
0 Point (43)
•
100% = 20 mA (36)
0% = 4 mA (37)