Etac Molift Mover 300 L Manuale d'uso - Pagina 18

Sfoglia online o scarica il pdf Manuale d'uso per Aiuto alla mobilità Etac Molift Mover 300 L. Etac Molift Mover 300 L 20.

Molift Smart 300 L / www.etac.com

Cleaning and disinfection

Clean on a regular basis. Clean surfaces with
a damp cloth using an appropriate pH-neutral
detergent. Do not use solvents or strong liquids,
this may damage surfaces on the lifter. For
disinfection when needed; use isopropyl alcohol.
Avoid abrasive cleaning products. Remove hair
and pile from the castors and verify that the
castors rotates freely. Clean contact and hand
control with isopropyl alcohol to remove grease
and dirt. Check emergency stop and emergency
lowering after cleaning. The lift should not be
exposed to running water.
Make sure not to damage or
remove labels when cleaning.

Periodic inspection scope

Periodic inspection is a visual examination (particu-
larly of the lifter's load bearing structure and lifting
mechanism with attachments, brakes, controls,
safety devices and person-support devices) accord-
ing to Periodic inspection Report for Molift Mover
300 L. This can be found on www.etac.com.
If there is any problems with the
lifter that could jeopardize some-
ones safety, the lifter shall imme-
diately be taken out of service and
marked "out of order". Do not use
the lifter until it is repaired.
Periodic inspection shall be performed annually or
more frequently if required by local requirements.
The inspection must be performed by service
personnel authorized by Etac.
Contact Etac at [email protected] for training and
authorization or recommendation of an approved
service partner.
When performing a periodic inspection, the
inspector shall fill out the inspection report
for Molift Mover 300 L. The reports should be

Maintenance

Reconditioning

Follow cleaning and assembly procedure, complete
periodic inspection and use checklist after assem-
bly to recondition the lifter.

Recycling

Refer to "Recycling instructions" on how to
properly dispose of the product. "Recycling
instructions" can be found on www.etac.com.

Spare Parts

A list of spare parts is available on request.

Periodic inspection

retained by the person(s) responsible for servicing
the lifter. If the inspection reveals defects and
damages, the owner shall be notified.
When periodic inspection is completed the
inspector shall mark the lifter with a sticker on
the control label showing the date when periodic
inspection is performed and a label (with 6-digit
ID from Etac or distributor, if available) showing
who has performed the periodic control. The
control label can be found on the right side of the
battery holder, and this will then indicate the date
when next inspection is due.
18
Mark label
with month
and year of
inspection