DOEATOOW HFS-10 Manuale operativo - Pagina 7

Sfoglia online o scarica il pdf Manuale operativo per Sensori di sicurezza DOEATOOW HFS-10. DOEATOOW HFS-10 10.

DOEATOOW HFS-10 Manuale operativo
5. History Records
All �ming measurement data is stored in the "Records"
interface.
The max storage of the device is ten pieces of �ming
measurement data.
If the device has stored ten historical records, the newly
created record will automa�cally overwrite the oldest one.
6. Data Clearing
Click the "Delete" menu and enter into the interface.
"Delete Current Data" is clearing the average value and
accumula�ng �me on the measurement page. "Delete
History Data" is clearing all data in the "Records" interface.
A�en�on: The Records cannot be recovered once they've
been cleared. When "
" shows on the upper screen, it
means the �ming measurement is in progress. At this �me,
unable to access the "Delete" interface.
Calibrate factors: Select "Calibrate Factor" and double click
it, then it will pop out "Password: 0000" on the bo�om of
the screen. Click
to adjust the password to "0018". Press
, then click
or
modify the CF value. Click
again
to save the changes. It is used to calibrate the error(20uSv -
1uSv).
Note: Default Cf=1.00
7. Alarm Limit Se�ng
Se�ng range of real-�me dosage alarm limit value:
0.1uSv-999uSv
Se�ng range of accumulated dosage alarm limit value:
1uSv-9999uSv
The unit will alarm when the reading exceeds the limit
value.