Cleaning

WARNING
• Prior to cleaning, make sure machine is turned off and unplugged.
• To ensure proper hygiene, collection container, inline filter, suction tubing, and intermediate tubing must only be used by the same patient.
Do not use these accessories across multiple patients. Do not attempt to sterilize these accessories for use across multiple patients. Failure to abide
by these warnings may result in infection or cross contamination between patients.
• Any parts or the device or accessories that come into contact with bodily secretions must be cleaned after each suction operation. All single-use
accessories, such as the suction catheter, must be disposed of according to local regulations, institution policy (as applicable) and the instructions
of a certified healthcare professional.
• Wear disposable gloves during cleaning procedures.
Cleaning the surface of the unit
• Clean the surface of the unit after each use. Wipe machine clean with a damp cloth.
• Do not clean machine with powder-type cleaning agents or soap.
• Do not clean machine in water.
WARNING: If the inside of the unit comes into direct contact with liquids, the aspirator must be sent to an authorized service center for
examination and repair.
ATTENTION: Slight discoloration of plastic housing or the stickers on the housing may occur over time. This will not impact the performance or
function of the device.
Cleaning reusable accessories
• Clean the collection container after each use. Empty contents and wash all components using a mixture of 3 parts white vinegar 1 part warm tap
water.
• Clean intermediate and suction tubing after each use. Disconnect tubing from machine and clean by rinsing inside and outside of tubing with hot
tap water. Air dry.
• Clean the collection container coaster after each use. Rinse with hot tap water and air dry.
WARNING
• To ensure proper hygiene, collection container, inline filter, suction tubing, and intermediate tubing must only be used by the same patient.
Do not use these accessories across multiple patients. Do not attempt to sterilize these accessories for use across multiple patients. Failure to abide
by these warnings may result in infection or cross contamination between patients.
• Fluid collected in the collection container must be disposed of according to local regulations, institution policy (as applicable), and the instructions
of a certified healthcare professional.
CAUTION: Do not immerse inline filter in liquid. Doing so will damage the filter. If the fibrous material inside the inline filter is wet, replace
immediately.
Maintenance & Repair
• Replace the inline filter if the fibrous material in the filter is wet or in the case of a collection container overflow. With the vacuum regulator fully
closed and the tubing disconnected from the "Fluid Side" of the filter (filter open to atmosphere), a vacuum reading in excess of 50 mmHg on the
gauge indicates the filter should be replaced. The filter must be replaced in the event fluids have been in contact with the filter, such as in a collection
container overflow.
• Do not disassemble the unit; there are no user-serviceable components inside the device. If repair or servicing is required, please send unit to an
authorized service center. Disassembly or tampering of the device will damage the device and void the warranty.
Warranty Length
The AS-1001D suction pump is warranted to be free from defective workmanship and materials for one year from the date of purchase. Within the
warranty period, any defective part(s) will be repaired or replaced at manufacturer's option. Manufacturer is not liable, and warranty is voided, for any
damage or consequential damage resulting from incorrect operation of the unit, inappropriate usage of the unit, disassembly of the unit, or use
by unauthorized or untrained persons.
Copyright © 2019 ANEST IWATA SPARMAX Co., Ltd. All rights reserved.