AKRUS AK 470 rk Manuale d'uso - Pagina 10
Sfoglia online o scarica il pdf Manuale d'uso per Arredamento per interni AKRUS AK 470 rk. AKRUS AK 470 rk 12. Patient examination stool
1.5
User instructions for safe operation
Before using this medical device, please carefully read and observe the safety instructions
o Correct use and operation of the table is absolutely necessary for its safe function.
o Check power specifications on label and ensure compliance with local power supply.
o Connect AK 470 only to power sockets equipped with ground circuit. Any break of ground
circuit (inside or outside of electrical equipment) may cause hazards to user or
equipment. Never break ground circuit on purpose.
o Connection directly to the examination unit by a trained service technician only. In this
case please observe a max. leakage current of 0,005 mA.
o Do not pull or tug on power cord.
o Do not operate the AK 470 series in environments exposed to hazardous materials. Do
not operate in the presence of flammable anaesthetics and volatile solvents such as
alcohol, benzine or similar
o Do not set up the AK 470 series in humid or damp rooms. Avoid areas of sprinkling,
dripping or splashing water.
o The AK 470 series should be operated by trained and authorized personnel only. Use
other than designed use falls within the sole responsibility of operator.
o Assembly, service or modifications by authorized and trained staff only.
o The AK 470 series may be opened only by trained and authorized technicians. Disconnect
power cord before opening equipment or changing fuses.
o When cleaning, make sure that no cleaning agent or water enters into the electric control
box of the AK 470 series
1.6
Intended use
The medical device is intended for use as a patient chair. Any other use of the product is not
permitted.
285g001e.docx
and recommendations listed in this manual
agents.
2022-02-03
page 4 of 6