医療機器 Ambu aScope 3のPDF クイックマニュアルをオンラインで閲覧またはダウンロードできます。Ambu aScope 3 2 ページ。 Intubation and single lung ventilation
Ambu aScope 3 にも: クイックマニュアル (2 ページ)

Ambu aScope 3 クイックマニュアル
Ambu
aScope
®
- Quick Guide for intubation and single lung ventilation
1.
Functions
The aScope 3 and aScope 3 Slim are the perfect choice
for all types of intubation as they are ready to use
whenever you need them. The instantly accessible
single-use and ready-to-go concept simplify flexible
intubation procedures and improve patient safety.
aScope 3 and aScope 3 Slim are single-use sterile
flexible videoscopes with a 2.2 mm or 1.2 mm working
channel respectively.
For intubation the aScope 3 can be used with ET tubes size
6 and up and the aScope 3 Slim with ET tubes size 5 and up.
For single lung ventilation aScope 3 Slim can be passed
through DLT's of FR 37 or greater or BB of sizes 5-9.
Warning: Read Instructions For Use before using the aScope
2.
Connecting the system
1. & 2. Connect aScope 3 or aScope 3 Slim to the
aView by plugging the aScope 3 connector
- white with blue arrow - into the corresponding
blue female connector on the side of aView.
3.
Preparation for use
Ensure that the selected ET tube or DLT/BB is compatible with
the aScope 3 or aScope 3 Slim.
Lubricate the insertion cord with a medical grade lubricant to
ensure the lowest possible friction. Do not get lubrication or any
other contaminant on the tip of the aScope 3/aScope 3 Slim.
Attach the suction tube to the suction connector and ensure
that the suction works when pressing the suction button
Fluids can be instilled through the working channel by inserting a
syringe into the working channel port at the top of the aScope 3.
When using a Luer Lock syringe, use the included introducer.
If you are using a preformed ET tube unbend it before inserting
the cord of the aScope 3 through it.
Antifogging agent is not required with this device.
Confirm that the bending section functions smoothly and
correctly by carefully sliding the bending control lever forward
and backward in each direction.
3 and Ambu
3 or aScope
3. Push the ON/OFF
button on the aView to
start up the system.
aScope
3 Slim
®
aScope 3
Insertion cord diameter 5.0 mm
Working channel 2.2 mm
aScope 3 Slim
Insertion cord diameter 3.8 mm
Working channel 1.2 mm
Bending angle 130º/130º
3 Slim and aView.
4. The real time image
will appear on aView
after only a few seconds
and the aScope 3 system
is ready to use
Confirm that the bending section returns to a straight position.
Connect the aScope 3/aScope 3 Slim to the aView and verify that
a live video image appears on the screen.
Always have an additional aScope at hand as a precaution.
Bending angle 150º/130º
Turn the page