4.
Preparation for use
The aView
TM
can be placed on a flat surface or secured to an I.V. pole using the
bracket supplied. The bracket also has a hook that facilitates easy, accessible
storage of the aScope
TM
3 via the hole in the top of the aScope
Most users find the default contrast and brightness settings to be optimal but
these can be adjusted if required.
Consider dimming the room lights to see the image from the aScope
Check the battery indicator on the aView
for the procedure charge the monitor by plugging it into the mains.
5.
Icons
Real time image will
appear after only a
few seconds.
After appr. 45 seconds
the user Interface is
ready displaying the
icons for battery status,
brightness/contrast
control, device usage
time as well as the as
snapshots and
recording of videos.
6.
Insertion of the aScope
The tube connection can be used
to fixate the selected ET tube during
intubation.
Keep the cord of the aScope
straight as possible by holding the
handle of the aScope
3 high. This
TM
increases the range of movement of
the tip and allows you more control.
Advance the aScope
3 slowly,
TM
allowing the light intensity of the
camera to re–adjust.
7.
Removal procedure
Slowly withdraw the aScope
The distal tip must be in a neutral and non–deflected position.
After use disconnect the aScope
accordance with local guidelines for collection of infected medical devices
with electronic components.
Wipe down the aView
as per the instructions in the Instructions For Use.
TM
Ambu A/S
Baltorpbakken 13 • DK-2750 Ballerup • Denmark
T +45 72 25 20 00 • F +45 72 25 20 50
www.ambu.com
TM
. If the time remaining is not sufficient
Start–up picture
When the aView
turned on the ON/OFF
button will light up
green. During charging
the ON/OFF button will
light orange.
3
TM
If the image is affected by blood or
secretion you can try to clear the
secretion by flicking the bending lever
up and down or dabbing the camera
3 as
TM
tip gently against the mucosal wall to
clear the image.
If the above fails, remove the aScope
and rub the lens clean with either
gauze or an alcohol swap. Do not be
afraid to exert a little pressure when
doing this. Proceed as above.
while observing the image on the monitor.
TM
from the aView
and dispose of it in
TM
TM
TM
3 pouch.
TM
3 more clearly.
Max. battery status
of the aView
TM
TM
is
Battery is charging
If the battery is fully charged and still
connected to a charger the battery
icon changes to:
TM
Min. batte ry
The icon remains white until one block is
status of the
left, after which it turns red.
aView
TM
When remaining battery capacity is 10%
the red battery icon starts flashing.
Charging is shown with blocks flashing.
Battery current
Current capacity is shown with non-
capacity
flashing blocks.