Checking the Electronics
Using the frequency oscillator and the cable that is supplied, connect it to the
electronic readout device, the same as you would a turbine flowmeter cable
assembly. Once the frequency oscillator is connected, if no output is displayed on
the electronic readout device, check to ensure that the power is on, the fuses are
good, and the correct K-factor is dialed/programmed into the electronic readout
device. After verification of these items, if there is still no output displayed and/or
registered, then the manufacturer of the electronic readout device should be
consulted.
When
the
frequency
connected and an output is displayed on
the
electronic
indicates that the turbine flowmeter cable
assembly is defective and should be
replaced. If no output is displayed, the
turbine flowmeter cable assembly is good
and the problem lies within the electronic
readout device.
14
oscillator
readout
device,
is
this