計測機器 CoaguSense RPO Assure PT CareのPDF クイック・リファレンス・マニュアルをオンラインで閲覧またはダウンロードできます。CoaguSense RPO Assure PT Care 2 ページ。 Prothrombin time (pt)/inr monitoring system
Assure® PT Care
Prothrombin Time (PT)/INR
Monitoring System
Quick Reference Guide
Getting Started and Setting Up the Meter
1. One Low and one High Control Strip should be run with each new shipment/lot of strips
received. Run controls if this has not been done yet. Controls are not required for
calibration.
2. Gather test supplies such as test strips, alcohol pads, lancets and sample transfer tubes.
3. Warm the patient's hands to increase circulation by washing the patient's hands with
warm soapy water.
4. Wipe the finger with alcohol. Dry patient's hands completely as alcohol can affect results.
Activity
1.
Turn the meter on by pressing and holding the POWER
button on the right side of the meter. Enter/Scan login
credentials (if required by facility). Press TEST.
2.
Confirm that the Patient Strip Lot # is correct by matching
with pouch. If blank or incorrect press CHANGE then align
and touch the NFC Tag rings to the target rings on the
meter surface. If the NFC tag is not available then manually
enter the Lot # and Barcode # from the patient test strip
pouch and press CONTINUE.
3.
Holding the round end, gently push the strip completely into
the meter. The meter warms the strip (for 25 seconds) to
operating temperature. The display shows a countdown in
seconds.
4.
At the end of countdown, the meter will beep once and
display "APPLY SAMPLE" for 2.5 minutes.
Intended Use -
The Assure PT Care Prothrombin Time (PT)/INR Monitoring System is an in vitro diagnostic device that provides
quantitative prothrombin time (PT) results, expressed in seconds and INR units. It uses fresh capillary whole blood. It is intended for use by
health care professionals at the point of care to monitor patients who are on warfarin-type (coumarin) anticoagulation therapy. The device is
not intended to be used for screening purposes.
Display/Visual Aid
PRO