- ページ 8

スイッチ Ametek Magnetrol SOLITELのPDF 設置および操作マニュアルをオンラインで閲覧またはダウンロードできます。Ametek Magnetrol SOLITEL 13 ページ。

8
5. Connect a digital multimeter between points MP5 on the
back of the module and ground connection (terminal 3 on
the electronic module) or the body of the vibrating probe.
See Figure 7 on page 7.
6. Adjust potentiometer P2 on the rear of the electronic mod-
ule to obtain a voltage between 0.4 volts and 1.0 volts. The
reading should be as close to 1.0 volt as possible.
7. Grasp the vibrating section of the sensor with your hand.
The vibration should stop and the relay will change state.
8. Remove your hand from the sensor. The relay will
change state.
9. The electronic module and probe are now matched.
3.2

Agency Approvals

3.2.1 CSA
Model
VXX-D1BX-XXX
These units have been evaluated to the
applicable UL and CS standards. CS is
accredited as a NRTL (National Recognized
Testing Laboratory) in the United States.
NRTL/C
These units have been tested to EN 50081-2
and EN 50082-2 and are in compliance with
the EMC Directive 89/336/EEC.
3.3

Replacement Parts

Item
1
2
3
4
5
6
7
8
9
Approval
Class I, Div. 2
Groups , B, C, & D
Class II, Div. 1
Groups E, F, & G
Type 4X
Description
Base
Cover
Cover o-ring
Electronics module:
110 V C
24 VDC
Bracket
Probe o-ring
Snap ring
Sleeve
Probe
56-601 Solitel Vibrating Rod Level Switch
Part Number
004-9189-001
004-9105-001
012-2101-345
Z30-9011-001
Z30-9011-004
005-6680-001
012-2408-001
010-5138-001
004-0111-001
Consult Factory