Refer to the 1200 Series Spectrophotometer Software User's Manual Version SS-
1.22 for detail.
Maintenance
Lamp Replacement
1. Turn off and unplug the instrument.
2. Remove the lamp access door at the back of the instrument by removing the four set
screws.
3. Unplug the lamp from the white connector. Insert the new lamp; pushing it in as far
as it will go. Replacement lamp: Item Number: S-1200-505 (6V 10W G4 type)
CAUTION: DO NOT HANDLE THE LAMP WITH BARE FINGERS. USE TISSUE OR
CLOTH WHEN HANDLING LAMP.
4. Turn on the instrument. Set the wavelength at 340 nm, insert a test tube or an empty
cuvette with cuvette holder, and blank the instrument. If the energy is low, adjust the
lamp by "pulling" or "pushing" it so that the light beam is focused on the entrance slit
of the monochromator. Since the lamp socket is pre-aligned, there will be minimum,
if any, adjustment required.
5. Re-install the lamp access door.
Wavelength Calibration Check
Normally the Cole Parmer 1200 Series spectrophotometer retains its wavelength
calibration indefinitely. However if the instrument receives a severe shock or is abused,
use the following methods to check wavelength calibration. Please note that this test
requires the Didymium filter, item number S-2100-116, or the Holmium Oxide filter, item
number S-2100-115.
In the filter method, the didymium filter has two distinct absorbance peaks at 529 nm
and 807 nm. The Holmium filter has a distinct peak at 361 nm. When the instrument is
calibrated properly you will find minimum Transmittance (maximum Absorbance) at the
range ± 2 nm from these peaks. Note that the specific Transmittance values are not
important as you are only looking for the wavelength where the minimum Transmittance
(maximum Absorbance) occurs.
Note: If you calibration filter has a certified peak/valley curve attached, please use the
peaks on the curve to verify the instrument.
Holmium Oxide Filter Method
1. Turn instrument on and allow it to warm up for 15 minutes.
2. Select the A mode.
3. Set the wavelength to 350 nm.
4. Make sure the Cuvette Adapter is empty and place it in the Sample Compartment.
Close the lid.
5. Set zero Absorbance by pressing the ∧ button. Wait a few seconds when the display
13