Wall attachment Mach 120/130
Mach LED 120/130
Note: Before doing any cleaning work turn off the lamp, disconnect from mains and make sure
that the lamp cannot be switched on again.
Please ensure that no disinfection liquid flows into the joints or lamp head.
The surface of the ceiling attachment can be easily kept clean by simply wiping with a wet cloth. You can
use conventional cleaning agents.
Caution:
In order to prevent any damage at plastic parts, do not use scouring agents or alkaline, acidic and
alcoholic cleaning agents.
For cleaning proceed as follows:
•
Draw plug from mains.
•
Wipe the surface of the ceiling attachment with a wet cloth. As a cleaning agent you can use water or
a soap solution (washing-up liquid).
Note: Before doing any maintenance work, turn off the lamp, disconnect from mains and make
sure, that the lamp cannot be switched on again.
The following maintenance work/tests has/have to be done every six months:
•
check on defects in paint work;
•
check on fissures at plastic parts;
•
check on deformation of the suspension.
The following maintenance work/tests has/have to be done once a year:
•
check the function;
•
electrical safety test;
•
check the suspension.
Remark:
Check and grease the security segment once a year.
59860002
6. Cleaning
7. Maintenance
7.1 Periodical maintenance work
Edition 11
Dr. Mach
Lamps and Engineering