- ページ 9
医療機器 Airssential Home AiroMax AI-MAXのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Airssential Home AiroMax AI-MAX 16 ページ。
Cleaning, Disinfection, Maintenance and Storage
Cleaning
Follow the procedure below to clean the accessories and compressor's housing.
1. Switch the compressor off and disconnect the power from the device.
2. Disconnect the air-tubing from the nebuliser bowl. Discard any residual medication
remaining in the nebuliser bowl. Rinse all parts of the nebuliser bowl and hence soak in
soapy, warm water for 5 minutes. After soaking, rinse all parts again with warm water.
Arrange the accessories on clean, absorbment towelling to dry before next usage.
Failure to clean nebulising accessories after use may decrease the effectiveness of
the treatment or lead to infection.
3. The compressor housing may be cleaned with a soft damp cloth. Do not use harsh or
abrasive cleaners, to avoid scratching the compressor's surface. DO NOT immerse
the compressor in any liquid. DO NOT rinse the compressor under running water.
Accessory Disinfection
Disinfect the nebuliser bowl before its initial use, and at the completion of each day's treatment,
or if the device has been unused for longer than 4 weeks. Follow the cleaning procedure
described above before disinfecting the nebuliser parts by using one of the methods below:
1. Disinfection is recommended using a medical disinfectant solution. Carefully follow
the disinfectant manufacturer's directions for use.
Alternatively, the accessories may be disinfected by soaking in a mixture of one-part
white vinegar with 3 parts of hot water, for 30 minutes.
2. After disinfecting, arrange the accessories on clean, absorbent towelling to thoroughly
dry before next usage.
Care of the Air-Tubing
1. After each use, check the air-tubing for condensation. If condensation is present, connect
only the air-tubing to the compressor and run the compressor to force air through the
tubing for up to 20 minutes, until the moisture has been removed by the airstream. Forcing
the moisture from the air-tubing in this manner alters the environmental conditions that
promote bacterial accumulation.
2. Do not soak or clean the air-tubing with water or any disinfecting solution. If the outside
of the air-tubing is dirty, wipe it clean with a soft, damp cloth and mild detergent. If the
inside of the air-tubing appears soiled or discoloured, it should be replaced.
AiroMax
Nebuliser System
®
9