計測機器 BD ALARIS 8015のPDF サービス速報をオンラインで閲覧またはダウンロードできます。BD ALARIS 8015 14 ページ。 Inspecting the pump module bezel
BD ALARIS 8015 にも: ユーザーマニュアル補遺 (8 ページ)

3. Remove the platen assembly and membrane frame assembly. Refer to the technical
service manual for disassembly instructions.
4. Inspect the bezel behind the membrane frame assembly for damage. If there is bezel
damage, the bezel assembly must be replaced. Otherwise, continue to Inspecting the
Bezel for Internal Damage on page 9.
Example of bezel damage behind the membrane frame shown in the figure below.
PN P00000360
Record number 1103-001-008-R-SD-INF, DIR 10000307230, Version 01 Release Date: 16AUG17
DOCO Use Only (Approver's List): 0207-001-001-S (Latest Version) to determine authorized approvers
DOCO Use Only (Record Format Template) 0204-011-005-R, DIR 10000153640, Ver 02, Release Date: 04APR16
Inspecting the Alaris™ Pump Module Bezel
DIR No.: 10000351665 Version: 00
Damage
Page 8 of 14