Inspecting the Bezel for Internal Damage
1. Close the pump module door and remove the rear case from the chassis assembly.
Refer to the technical service manual for disassembly instructions.
2. Inspect the bezel for damage. If there is bezel damage, the bezel assembly must be
replaced. Otherwise, continue to Inspecting the Bezel Posts on page 10.
Examples of bezel damage are shown in the figures below.
PN P00000360
Record number 1103-001-008-R-SD-INF, DIR 10000307230, Version 01 Release Date: 16AUG17
DOCO Use Only (Approver's List): 0207-001-001-S (Latest Version) to determine authorized approvers
DOCO Use Only (Record Format Template) 0204-011-005-R, DIR 10000153640, Ver 02, Release Date: 04APR16
Inspecting the Alaris™ Pump Module Bezel
DIR No.: 10000351665 Version: 00
Damage
Damage
Page 9 of 14