14. TROUBLESHOOTING (2)
PROBLEM
LCD shows low battery
symbol
LCD shows "Er 0"
LCD shows "Er 1"
LCD shows "Er 2"
LCD shows "Er 3"
LCD shows "Er 4"
LCD shows "Er 5"
LCD shows "Er 6"
LCD shows "Er 7"
LCD shows "Er 8"
LCD shows "Er A"
No response when you
press button or load
battery.
Speaking, arm or body movement,
angry, excited or nervous during
testing
Irregular heartbeat (arrhythmia)
POSSIBLE CAUSE
Low Battery
Pressure system is unstable
before measurement
Fail to detect systolic pressure
Fail to detect diastolic pressure
Pneumatic system blocked or cuff
is too tight during inflation
Pneumatic system leakage or
cuff is too loose during inflation
Cuff pressure above 300mmHg
More than 3 minutes with cuff
pressure above 15 mmHg
EEPROM accessing error
Device parameter checking error
Pressure sensor parameter error
Incorrect operation or strong
electromagnetic interference.
re-test.
Re-test when calm and without
speaking or moving during the
test
It is inappropriate for people with
serious arrhythmia to use this
Electronic Sphygmomanometer.
11/17
SOLUTION
Change the batteries
Don't move and try again.
Apply the cuff correctly and
try again
Measure again after five
minutes. If the monitor is
still abnormal, please
contact the local distributor
or the factory.
Take out batteries for five
minutes, and then reinstall
all batteries.
Template version:1