B. Automatic Mode allows the user to load up to 32 sample tubes into the carousel and
program the instrument to automatically run each tube in sequence. Automatic Mode
is the default setting for the CXP Software; no special button must be pressed to
enable Automatic Mode.
1. Adjust the Workspace Preferences, if needed, to produce the desired naming
and printing results (see "Naming LMD Files" and "Setting Printing
Options").
2. The Acquisition Manager window must be visible. If it is not, click View
and then Acquisition Manager.
3. If samples were being run in Single Tube mode (
click
4. Many people find it helpful to change the display of the Acquisition
Manager window. To do so, click View, then Customize Worklist Columns.
Uncheck Tube ID. Check Sample ID2. Adjust column size in the
Acquisition Manager window.
5. Click the Clear Worklist icon
6. For each sample being run, click on the Insert Test icon
7. In the Acquisition Manager window, enter what is desired in the Sample ID1
column. Hit Enter, and this text will be added to each test and cannot be
changed for each individual test.
8. Here is where organization is essential! In Sample ID2, enter the specimen
names. When entering the first sample name, after hitting Enter the software
applies this name to all of the tests. Simply use the down arrow key and type
the second sample name, then down arrow and type the third, etc. The
software should allow you to simply overwrite.
9. Lift the lid of the carousel and load sample tubes into the appropriate
numbered slots. Ensure that the sample name entered into the Acquisition
Manager worklist corresponds to the sample tube in the corresponding
numbered slot in the carousel.
10. Examine the carousel and identify which number it is (1 or 2) by the white
writing on the top knob. Type the number into the Carousel Number column
of test 1 of the Acquisition Manager window, then hit Enter.
11. To run the carousel, click the Run button
again to take it out of Single Tube mode.
clicked and shaded),
in the toolbar.
in the toolbar.
in the toolbar.