Running Tubes
A. Single Tube Mode allows the user to place one tube at a time onto the carousel. The
sample is run, then the instrument stops & returns the tube. The first tube must be
removed and then another tube can be loaded in the same position. All tubes are
placed in position 10 of the carousel, which can be accessed by the small Tube
Access door on the front of the carousel lid when the instrument is in Single Tube
mode. When in Single Tube mode, the lid of the carousel should not be opened.
1. Adjust the Workspace Preferences, if needed, to produce the desired naming
and printing results.
2. Click on the Single Tube button
3. Lift the Tube Access door, insert tube, and close door.
4. Click on the Run button
i. While running, voltages and compensation values can be altered from
the Detectors and Compensation tabs, respectively, on the
Cytometer Controls window, accessed by the button on the toolbar or
by clicking Cytometer then Cytometer Controls.
ii. If setting up a new protocol, it is advisable to alter the
voltages/compensation under Setup mode, which can be activated
through a checkbox in the Parameters tab of Cytometer Controls.
5. After starting the run, the software will display a series of messages at the
bottom of the screen: Awaiting Sample, Preparing Sample, Acquiring, and
Stopping. When complete, the tube will return to the Tube Access door and
the software will display the Awaiting Sample message.