ラボラトリー機器 Ankom XT15IのPDF サービス手順をオンラインで閲覧またはダウンロードできます。Ankom XT15I 5 ページ。 Extractor sight glass assembly

3.
With the front plate removed, observe the sight
glass assembly.
4.
Viewing from the back panel, cut the zip ties from
the Sight Glass Assembly's tri-fold tubing.
5.
Use a boxcutter to cut small slits in the tubing to
remove all three from their connections.
ANKOM Technology • 2052 O'Neil Road •
Sight Glass Assembly (X85)
XT15/XT15I
https://www.ankom.com/contact/technical-services
Zip ties
• Phone: 315-986-8090
#021
Service Procedure
Revised: 11/1/22
pg. 2 of 5