Ankom XT15I Procedura serwisowa - Strona 2

Przeglądaj online lub pobierz pdf Procedura serwisowa dla Sprzęt laboratoryjny Ankom XT15I. Ankom XT15I 5 stron. Extractor sight glass assembly

3.
With the front plate removed, observe the sight
glass assembly.
4.
Viewing from the back panel, cut the zip ties from
the Sight Glass Assembly's tri-fold tubing.
5.
Use a boxcutter to cut small slits in the tubing to
remove all three from their connections.
ANKOM Technology • 2052 O'Neil Road •
Sight Glass Assembly (X85)
XT15/XT15I
https://www.ankom.com/contact/technical-services
Zip ties
• Phone: 315-986-8090
#021
Service Procedure
Revised: 11/1/22
pg. 2 of 5