Figure 8. 529 5305 Secondary Remote Alarm
When making connections between the remote alarms, move the power switch to the off position (down) on the
529 5298 or 529 5300 remote alarm and disconnect the power. Wire the connections as shown in Figure 9; some
connections will be unnecessary if any of autoswitches 1-4 are not present. Strip the end of each wire about 1/8"
and insert the wire into the appropriate terminal. Tighten the terminal block screw using a small flat-tip screw-
driver.
Connect the power source and turn on the power to the 529 5298 or 529 5300 remote alarm. The power LED's on
the autoswitch, the secondary remote alarm, and the remote alarm should illuminate. Check the function of the
system by pressurizing all banks and allowing one side to depressurize at a time. When a bank is low (below the
alarm set point on the gauge), the LED will illuminate on the remote alarm, the secondary remote alarm, and the
autoswitch. The buzzer on all of the alarms will sound if no other alarm LED's are on. Silence the buzzer by
pressing the remote alarm's silence button.
Figure 9. Connections between two remote alarms
All subsequent 529 5305 remote alarms (up to 2 more) should be connected in series off of the first 529 5305
secondary remote alarm. Match terminal to terminal.