Command 5: Data Range Settings
{ 5, Channel Select, Data Select, Data Begin, Data End, Step, K (, FFT Samples) }
Cannot be used during sampling.
1. Channel Select
0
Highest priority data
1
Specify CH1.
2
Specify CH2.
3
Specify CH3.
4
Specify SONIC channel.
5
Specify Digital In
6
Specify recorded time data.
10 Specify mic.
Send sequence of the specified channel's sampled data takes priority.
2. Data Select
0
Raw data
1
d/dt
2
d
2
/dt
2
9
A/D value
10 FFT (real number part)
11 FFT (imaginary number part)
Send sequence of the specified data takes priority.
For calculation method and other information, see "3. Post-Processing" on page 10.
3. Data Begin
1–120000
Sends data specified by number.
4. Data End
1–120000
0 Last sample
Sends from data start number up to number of data specified by this number.
When 0, sends up to end of data.
5. Step
1– Number of steps (Specifying 2 sends ever other data item).
–1 Send using step equivalent to [Number of Sampled Data]
(rounded up).
FFT Samples
1 to 13
FFT calculation number of samples 2^n
* This parameter is ignored when "Data Select" is not "FFT".
– 15 –
[Specified Value K]