Casio E-Con EA-200 기술 참조 - 페이지 15
{카테고리_이름} Casio E-Con EA-200에 대한 기술 참조을 온라인으로 검색하거나 PDF를 다운로드하세요. Casio E-Con EA-200 20 페이지. Casio ea-200 calculator-tools: user guide
Casio E-Con EA-200에 대해서도 마찬가지입니다: 제품 카탈로그 (16 페이지)
Command 5: Data Range Settings
{ 5, Channel Select, Data Select, Data Begin, Data End, Step, K (, FFT Samples) }
Cannot be used during sampling.
1. Channel Select
0
Highest priority data
1
Specify CH1.
2
Specify CH2.
3
Specify CH3.
4
Specify SONIC channel.
5
Specify Digital In
6
Specify recorded time data.
10 Specify mic.
Send sequence of the specified channel's sampled data takes priority.
2. Data Select
0
Raw data
1
d/dt
2
d
2
/dt
2
9
A/D value
10 FFT (real number part)
11 FFT (imaginary number part)
Send sequence of the specified data takes priority.
For calculation method and other information, see "3. Post-Processing" on page 10.
3. Data Begin
1–120000
Sends data specified by number.
4. Data End
1–120000
0 Last sample
Sends from data start number up to number of data specified by this number.
When 0, sends up to end of data.
5. Step
1– Number of steps (Specifying 2 sends ever other data item).
–1 Send using step equivalent to [Number of Sampled Data]
(rounded up).
FFT Samples
1 to 13
FFT calculation number of samples 2^n
* This parameter is ignored when "Data Select" is not "FFT".
– 15 –
[Specified Value K]