- ページ 8

医療機器 CONTEC MEDICAL SYSTEMS 2.3.10.00002のPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。CONTEC MEDICAL SYSTEMS 2.3.10.00002 14 ページ。 Pulse oximeter probe

CONTEC MEDICAL SYSTEMS CO.,LTD
Product name:
Pulse Oximeter Probe
Product model:
GSE0002(
Scope of application:
This product is used to match CONTEC Patient Monitor , Electronic Sphygmomanometer,Pulse Oximeter collect and transmit the
SpO
signal from patient with continuance and no trauma. It is inapplicable to monitor the weak perfusion moving state and monitor
2
for long, so check the measuring position or change for another position per 4 hours.
Taboo disease:
Don't fix the product on the position with tissue injury. It is inapplicable for the patient or users allergic to PVC、TPU、TPE、ABS plastic.

Product performance:

1) The range of SpO
measurement: 70%~100%;
2
Accuracy: 70~100%:±2%; Below 70%: unspecified.
2) The range of pulse measurement: 30~250bpm;
Accuracy: ±2bpm or ±2%(select larger).
3) Optical Sensor:
Red light (wavelength is 650~670nm, 6.65mW)
Infrared (wavelength is 880~910nm, 6.75mW)
Main configuration:
Consisting of plug, cable and probe.
Power supply requirement:
Sphygmomanometer,
Pulse Oximeter .which are applicable to the requirements of IEC60601-1.
Directions for use:
Note:
This product is type BF applied part;
Sketch map
1) As Figure 1, the pulse oximeter probe of different types is applied to different crowds.
2)
Select proper probe and put recommendatory placement according to Figure 1.
3)
Arrange the cable along the back of hand when place the pulse oximeter probe.
4) Connect
pulse oximeter probe with pulse oximeter ,Electronic Sphygmomanometer,or patient monitor and check if the operating
procedure accords with the procedure introduced in user manual.
5)
Pulse Oximeter Probe GSE0002 (2.3.10.00002) needs the help of the FST0001 pulse Oximeter probe extension cable to be
connected into the jack of the Pulse Oximeter CMS60D, CMS70A for normal use. Connect the FST0001 pulse Oximeter probe extension
cable into the jack of the Pulse Oximeter CMS60D, CMS70A, then connect pulse oximeter probe to the other end of the FST0001 pulse
Oximeter probe extension cable. Plug the sensor connector firmly into the FST0001 blood oximeter extension cable.
Pulse Oximeter Probe GSE0002 (2.3.10.00002) needs the help of the FST0004 pulse Oximeter probe extension cable to be connected
into the jack of the Electronic Sphygmomanometer CONTEC08A for normal use. Connect the FST0004 pulse Oximeter probe extension
Pulse Oximeter Probe user manual
2.3.10.00002 &
2.3.10.00009), GSE0003(2.3.10.00006), GSE0004(2.3.10.00007), GSE0005(2.3.10.00008)
The special power is supplied by from the equipments of CONTEC Patient Monitor,
Model explanation
Adult disposable
Pediatric disposable
Infant disposable
(2.3.10.00002 & 2.3.10.00009)
Neonatal disposable
GSE0003
(2.3.10.00006)
sensor
GSE0004
Weight10~50kg
(2.3.10.00007)
sensor
GSE0005
(2.3.10.00008)
sensor
GSE0002
sensor
/weight>40kg adult
Figure 1
Applied crowds
Weight>30Kg
Recommendatory
adult
placement: forefinger
Recommendatory
Pediatric
placement: forefinger
Weight3~20kg
Recommendatory
Infant
placement: big toe
Weight<3kg
Recommendatory
neonate
placement: neonate's
adult's forefinger
Electronic
Placement
foot,
1