CONTEC MEDICAL SYSTEMS CO.,LTD
cable into the jack of the Electronic Sphygmomanometer CONTEC08A, then connect pulse oximeter probe to the other end of the
FST0004 pulse Oximeter probe extension cable. Plug the sensor connector firmly into the FST0004 blood oximete extension cable.
Pulse Oximeter Probe GSE0002 (2.3.10.00002) needs the help of the FST0014 pulse Oximeter probe extension cable to be connected
into the jack of the new Patient Monitor CMS8000 for normal use. Connect the FST0014 pulse Oximeter probe extension cable into the
jack of the Patient Monitor CMS8000, then connect pulse oximeter probe to the other end of the FST0014 pulse Oximeter probe
extension cable. Plug the sensor connector firmly into the FST0014 pulse oximeter extension cable.
Notice items:
1) pulse oximeter probe placement, the position without ductus arteriosus, BP cuff and vein input pipe is top-priority.
If the pulse oximeter probe can't monitor the state of pulsation, it shows that the position of probe is improper, or the position is too
2)
thick, too thin or having too deep pigment to reach a proper translucidus effect. If above things has happened, place the probe again or
select probe of other type.
This pulse oximeter probe should be applied to the special medical equipment. Operator is responsible to check the compatibility
3)
Incompatible fittings or device will influence the measuring result.
The disposal of scrap instrument and its accessories and packing (including battery, plastic bags, foams and paper boxes) should
4)
follow the local laws and regulations.
Maintenance/cleaning/disinfection:
1) Check if the product is undamaged and clean before using.
2)
This product is not allow to use disinfection liquid for disinfection, this probe belong to one-off products.
Note: Don't immerse the product in the liquid, and don't expose it under the strong ultra-violet radiation
Service life:
Suggest this product use only once, don't use again.
Environment requirements:
Transport and storage
1)
Temperature: -10℃~+40℃
2)
Humidity: less than 80%
3)
Pressure:86kPa~106kPa
Operating
1)
Temperature: 10℃~+40℃
2)
Humidity: 30% ~ 75%
3)
Pressure: 700hPa~1060Pa
Statement:
1)
pulse oximeter probe needs special precautions regarding EMC and needs to be installed and put into service according to the
EMC information provided in User Manual and test report.
2)
Portable and mobile RF communications equipment can affect pulse oximeter probe.
Warning:
1) The use of cables other than those specified, with the exception of cables sold by CONTEC as replacement parts for internal
components, may result in increased emissions or decreased immunity of pulse oximeter probe.
pulse oximeter probe should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary,
2)
the pulse oximeter probe should be observed to verify normal operation in the configuration in which it will be used.
3) Improper usage can result in inaccurate measurement.
Using it under too strong light will cause inaccurate measurement, in case of that, please set a opaque stuff around the probe to cut
4)
light off.
You should move the probe to other position per 4 hours at least. Because the state of local skin can influence the ability of skin to
5)
enduring probe, it is necessary to replace the position of probe according to the state of patient. Please do that when skin integrity
changes.
6)
The dyestuff in blood vessel cab cause the inaccurate measurement.
The performance of pulse oximeter probe is influenced by movement easily, so it is not suitable for active patient to use it.
7)
8) Don't fix the probe with belt or bundle it tightly, because the vein pulsation can cause inaccurate SpO
9)
Same as other medical equipment, the cable should be set properly to avoid enlacing or asphyxiate patient.
10) Don't use it in the process of MRI scan, because the conductor current may burn the skin of patient, moreover, the probe will
measurement.