OPERATION AND MEASUREMENT
Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying
or recording a millivolt signal (an input measurement range of approximately 0 to 12.5 mV is required to cover the
entire range of red and far-red photon flux density from the sun). In order to maximize measurement resolution
and signal-to-noise ratio, the input range of the measurement device should closely match the output range of the
quantum sensor. DO NOT connect the sensor to a power source. The sensor is self-powered and applying voltage
will damage the sensor.
Wiring for S2-131
White: Positive (signal from far-red sensor)
Blue: Negative (signal from red sensor)
Yellow: Positive (signal from red sensor)
Clear: Shield
Black: Negative (signal from far-red sensor)