- ページ 9
計測機器 Apogee Instruments SIF-121のPDF オーナーズマニュアルをオンラインで閲覧またはダウンロードできます。Apogee Instruments SIF-121 17 ページ。 Infrared radiometer
Apogee Instruments SIF-121 にも: オーナーズマニュアル (19 ページ)
Field of View
The field of view (FOV) is reported as the half-angle of the apex of the cone formed by the target surface (cone
base) and the detector (cone apex), as shown below, where the target is defined as a circle from which 98 % of the
radiation detected by the radiometer is emitted.
Sensor FOV, distance to target, and sensor mounting angle in relation to the target will determine target area.
Different mounting geometries (distance and angle combinations) produce different target shapes and areas, as
shown below.
A simple FOV calculator for determining target dimensions based on infrared radiometer model, mounting height,
and mounting angle, is available on the Apogee website:
http://www.apogeeinstruments.com/using-your-apogee-
instruments-infrared-radiometer/.