- ページ 7

体重計 ADE M703600-01のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。ADE M703600-01 12 ページ。 Electronic lifter scales

ADE M703600-01 取扱説明書
Once Backlight is set on, scale never turns off backlight,
independent from stability.
Press
to exit the Settings Mode.
Symptom
blank display by pressing
On/Off
no action taken after TARE or
HOLD is activated
unstable readings
Scale does not show zero
without load
substantial error in weight
result

Cleaning:

For cleaning just use a damp cloth or common disinfection products.
Do not use any aggressive cleaning fluids.
Warranty – Liability:
ADE will repair or replace this product within a period of 1 year from date of delivery applying to
defects occurred due to poor material or workmanship (presentation of invoice is necessary). All
moveable parts – batteries, cables, mains units, rechargeable batteries etc. – are excluded. Warranty
does not include natural wear, damages caused by incorrect or negligent handling, overuse or
unsuitable or improper storage. The cost of transport will be borne by the customer should the
equipment be located anywhere other then the customer's premises. In the event of transport
damage, claims under warranty can be accepted subject to using the complete original packaging and
the equipment secured the way it was packed and secured originally. All the packaging should
therefore be retained.
Any warranty claim will be rejected if the equipment is opened by persons not authorized expressively
by ADE to do so.
To make sure the scale is fully reliable at any time do check the scale in regular intervals by using
tested weights.
M703600+M703600-01-210817-Rev008-UM-en
Trouble-shooting & Cleaning
Possible reasons
discharged battery
defective battery
defective ON/OFF key
defective TARE or HOLD key
scale in motion
scale is damped
dust on PCB board
unstable power system
load-cell stressed too much or
for too long
Scale has been loaded before
zero setting process has been
finished
re-calibration needed
improper unit
Possible actions
replace battery
press ON/OFF key for long
clean TARE or HOLD key
stabilize the load and scale
dry the scale
clean PCB board
increase warm-up time
hang the scale during storage
keep the scale unloaded and
start procedure again and wait
for zero setting before loading
re-calibrate the scale
switch to proper unit
7