- ページ 14

産業機器 dymax T17240のPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。dymax T17240 20 ページ。 Digital syringe dispenser

dymax T17240 ユーザーマニュアル
14
SD-100 Digital Syringe Dispenser User Guide
2.
The display will show Ld 1 (Load from Memory #1).
3.
Press the MEM key again to change from Ld 1 to Sv 1 (Save to Memory #1).
4.
Press the ▲ key 8 times to get to memory address 9.
5.
Press the EXIT key to complete the save.
To save your current cycle mode settings, switch the mode to cycle and then continue
with the above procedure.

Loading from Memory

After saving your cycle and timer mode settings to the SD-100 memory, you can recall
them to be used again. Both the cycle mode and timer mode need to be loaded
separately by switching to the mode you would like to load.
For example: To load the cycle time that was saved in Memory 9 on the previous page,
follow the steps below.
1.
Switch to cycle mode then press and hold the MEM (Memory) key for 2 seconds
until the unit beeps.
2.
The display will show Ld 1 (Load from Memory #1).
3.
Press the ▲ key 8 times to get to memory address 9.
4.
Press the EXIT button to complete the load. Program 9 cycle mode settings are now
in effect. The display should show 5 seconds.
To load the timer mode settings, switch to timer mode then follow the same directions.