Features
The 7829 Visconic offers features that are suited to specific applications and industry needs.
Process Requirement
Dynamic viscosity
Kinematic viscosity
Temperature measurement
Integrated density measurement
Referred kinematic viscosity
Referred density measurement
Sensor type
Long-stem version available
Mounting arrangement
Transmitter version
mA outputs
Measurable liquid types
Principle markets
No Field Calibration Required
The 7829 viscosity meter is factory-calibrated and
no field calibration is required. The calibration is
traceable to UK National Standards through the
Micro Motion onsite accredited laboratory.
Integral Electronics for Signal Processing
A configured microprocessor-based electronic
module provides the full signal conditioning,
calculation, and diagnostic facilities within these
meters. The 7829 meter does not require remote
electronics for signal processing. This meter
simultaneously measures real-time viscosity,
density, and temperature; and, the integrally
mounted electronics can be configured to output a
number of user-defined parameters.
User-Configurable 4–20 mA Outputs
Two 4–20 mA outputs are available and configurable
to output parameters such as dynamic and
kinematic viscosity, density, temperature and other
calculated measurements such as base viscosity,
°API, °Brix, % solids, % mass, % volume and
specific gravity.
Any of the output parameters can be used to drive
the two 4–20 mA outputs, enabling them to be used
as the process variable in control applications
without the need for additional processing
electronics.
4
Micro Motion 7829 Visconic Viscosity Meter
7829 Visconic
Bypass / Tank / Inline
Two 4–20 mA (both configurable)
Refined products and crude oil
Following are the default settings of the two
4–20 mA outputs.
mA Output 1
Viscosity (cSt)
Code
4 mA
H
0
J
0
E
0
M
0
K
0
F
0
G
0
Z
RS-485/Modbus Communications Interface
All measurements are available digitally through the
built-in RS-485/Modbus communications interface,
for integration into plant data systems. You can use
the ADView or ProLink II (v2.9 or later) software
tools to configure or diagnose a meter via the
RS-485/Modbus interface. See "ADView Software
Tool Features" and "ProLink II Software Tool
Features" for more information.
Fork
Industrial
mA Output 2
Temp (°C)
20 mA
4 mA
25
0
50
0
100
0
200
0
500
0
1000
0
12500 cP
0
Custom
20 mA
150
150
150
150
150
150
150