- ページ 7
コントロールユニット A-M Systems 2200のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。A-M Systems 2200 19 ページ。 Analog stimulus isolator
NOTE: If the Model 2200 is beeping, then the 9V battery needs to be replaced.
If for some reason it is necessary to store the 2200 for more than a few months, it is
recommended that the unit be kept at a low temperature(5-10°C). Before storing for longer
than a year, it is advisable to remove the batteries from the unit; store the batteries separately
Control Port
Digital control is performed via the subminiature DB-9 connector on the instrument.
Signal Description
The output signal is turned on and off via the Gate line. The default (open circuit) condition is
a logical "1", allowing unimpeded signal transmission from input to output.
Output errors may be detected via the Error line. Normally at a logical "0", this line goes high
for at least 500ms if the output amplifier cannot drive the load satisfactorily or if the output
range were exceeded, such as trying to obtain 200V from the unit.
Since it is possible that either or both of the internal batteries will be too low to provide a
readable Error signal, it is strongly recommended that a periodic battery check be done via the
front panel button in order to ensure proper operation.
Pinout
The pins are connected as follows. Note that while this is mechanically compatible with a 9-
pin RS-232 connector, it is not electrically compatible. The 2200 is not smart enough to
understand serial communications; and for some, the slow, erratic speed of serial
communications would compromise performance
A-M Systems
Figure 1. Bottom view of case: screw locations
PO BOX 850, Carlsborg, WA 98324 USA
Telephone: 800-426-1306 * 360-683-8300 * FAX: 360-683-3525
E-mail: [email protected] * Website: http://www.a-msystems.com